XSLMS High Speed Spectral LED Measurement System
High-speed analysis of LED and SSL Lamps and Assemblies FEATURES:
- Wide Dynamic Range Allows a Single Sphere to Measure a Wide Range of Light Levels
- NIST Traceable Standards for In-house Recalibration
- Spectral Results in Milliseconds
- Comprehensive Light Measurement Software
- Total Spectral Flux - Luminous Flux - Radiant Flux - Chromaticity - CCT - CRI - Peak Wavelength - Dominant Wavelength - Spectraflect® Interior Sphere Coating
- CIE Recommended Geometry
MEASURE:
- Emission Spectrum
- Reflectance Spectrum
- Transmittance/Absorption
The XSLMS System combines Labsphere's intuitive software with high-speed hardware to simplify complex LED and SSL Measurements. Sphere diameters for the systems range from 10 inches to 3 meters and are available in either Labsphere's full or HalfMoon integrating sphere assemblies, with a choice between four high performance 9800 series CCD-based spectrometers. A variety of interchangeable light measurement accessories are available to customize your system to meet specific measurement requirements.
The integrating spheres are coated with Labsphere's Spectraflect® High Reflectance Coating for a nearly Lambertian sphere interior. The 9800 Series Spectrometers provide a wide dynamic range for a variety of total flux measurement needs and feature a 20% lower stray light effect compared to other spectrometers. Labsphere's intuitive Windows-based LightMtrX® Software guides users through testing, providing straightforward procedures and repeatable results to meet specific needs.
FOUR HIGH SPEED SPECTROMETER MODELS OFFER WIDE DYNAMIC RANGE AND FLEXIBILITY TO MEET YOUR EXACT MEASUREMENT NEEDS.
The XSLMS Series Systems are available in a choice of 4 spectrometer wavelength ranges to give you flexibility in choosing a system that’s just right for your application.
- Wide dynamic range covers a variety of total flux measurement needs
- Suitable for UV and quantum efficiency evaluation due to 1/5 lower stray light effect compared to conventional spectrometers
- Exposure time range 5 msec to 20 sec (option: 5 msec to 65 sec) covers various measurement needs, including extremely low light, high speed and production line
- High repeatability achieved with Quartz scramble fiber bundles
- Lightweight and compact design




XSLMS High Speed Light Measurement System
Ordering Information and Specifications
System Includes
|
10 inch Sphere System
|
20 inch Sphere System
|
40 inch Sphere System
|
65 inch Sphere System
|
76 inch Sphere System
|
Light Measurement Sphere
|
LMS-10
|
LMS-20
|
LMS-40
|
LMS-65
|
LMS-76
|
Spectrometer
|
Choose below
|
Choose below
|
Choose below
|
Choose below
|
Choose below
|
Power Supply, Lambda TDK
|
GEN-150-10
|
GEN-150-10 |
GEN-150-10 |
GEN-150-10 |
GEN-150-10 |
| Light Measurement Software, Required, Sold Separately |
XSLMS-SPEC
|
XSLMS-SPEC |
XSLMS-SPEC |
XSLMS-SPEC |
XSLMS-SPEC |
Lamp Standard
|
TOCS-10
|
TOCS-20
|
TOCS-40
|
TOCS-65
|
TOCS-76
|
Sphere Specifications
|
10 inch Diameter (25 cm) |
20 inch Diameter (50 cm) |
40 inch Diameter (1.02 m) |
65 inch Diameter (1.65 m) |
76 inch Diameter (1.93 m) |
Sphere Coating Reflectance:
|
98%
|
98% |
98% |
98% |
98% |
Radiometric Range:
|
100 W (max)
|
400 W (max)
|
1,500 W (max)
|
4,000 W (max)
|
5,000 W (max)
|
Minimum Lumen Level:
|
|
|
|
|
|
Illuminant A:
|
1.3 mlm
|
2.0 mlm
|
7.0 mlm
|
20 mlm
|
27 mlm
|
Red LED Range:
|
0.20 mlm
|
0.22 mlm
|
2.1 mlm
|
4.0 mlm
|
5.7 mlm
|
Green LED Range:
|
0.40 mlm
|
0.50 mlm
|
2.0 mlm
|
5.0 mlm
|
6.7 mlm
|
Blue LED Range:
|
0.13 mlm
|
0.17 mlm
|
0.83 mlm
|
1.7 mlm
|
2.3 mlm
|
Max Recommended DUT Dimension:
|
1 x 1 in (3 x 3 cm)
|
2 x 2 in (5 x 5 cm)
|
4 x 4 in (10 x 10 cm)
|
7 x 7 in (18 x 18 cm)
|
8 x 8 in (21 x 21 cm)
|
Sphere Dimension: (W x D x H)
|
16.7 x 11.9 x 15.5 in (42.4 x 30.2 x 39.4 cm)
|
28.5 x 23.7 x 29.4 in (72.4 x 60.2 x 74.7 cm)
|
48.9 x 42.3 x 68.5 in (1.24 x 1.07 x 1.74 m)
|
112.6 x 78.4 x 84.1 in (2.86 x 1.99 x 2.13 m)
|
116.9 x 88.2 x 91.4 in (2.97 x 2.24 x 2.32 cm)
|
2 pi External Port Dimension:
|
2 in (5.1 cm)
|
6 in (15.2 cm)
|
8 in (20.3 cm)
|
12 in (30.5 cm)
|
18 in (45.7 cm)
|
Lamp Standard Specifications
Lamp Standard
|
SCL-600
|
SCL-600
|
SCL-1400
|
SCL-1400
|
SCL-1400
|
Approximate Luminous Flux:
|
600 lm
|
600 lm
|
1400 lm
|
1400 lm
|
1400 lm
|
Lamp Current: (amps)
|
2.600
|
2.600
|
2.686
|
2.686
|
2.686
|
Rated Voltage: (volts)
|
12.8
|
12.8
|
28
|
28
|
28
|
Rated Life Base: (hours)
|
600
|
600
|
2000
|
2000
|
2000
| CDS-9800 Series Spectrometers (Choose your own)
CDS-9800-VIS Spectrometer (360 - 830 nm)
Detector: (*1)
|
TE Cooled 1024 x 122 CCD
|
Spectral Range:
|
360 - 830 nm |
Grating:
|
f= 85.8 mm 448 grooves/mm
|
Gate-Time:
|
5 msec - 20 msec (option: 5 msec to 65 sec)
|
ND Filter:
|
OD1-2
|
Data Point Interva:l
|
0.5 nm (1024 ch)
|
Resolution: (nm/ch)
|
1.0 nm (512 ch)
|
Wavelength Accuracy:
|
+ 0.3 nm
|
Stray Light: (B-Band for Illuminant A)
|
2.5 - 10E-4
|
Dynamic Range:
|
1000000:1
|
Linearity:
|
+ 0.8%
|
A/D Converter:
|
16 bit
|
A/D Rate:
|
1 MHz
|
PC Interface:
|
USB 2.0
|
Weight:
|
6 kg (18 lbs)
|
Dimensions: (W x H x D)
|
105 x 230 x 282 mm | CDS-9800-VIS-IR Spectrometer (360 - 1100 nm)
Detector: (*1)
|
TE Cooled 1024 x 122 CCD
|
Spectral Range:
|
360 - 1100 nm |
Grating:
|
f= 85.8 mm 303 grooves/mm
|
Gate-Time:
|
5 msec - 20 msec (option: 5 msec to 65 sec)
|
ND Filter:
|
OD1-2
|
Data Point Interval:
|
0.8 nm (1024 ch)
|
Resolution: (nm/ch)
|
1.6 nm (512 ch)
|
Wavelength Accuracy:
|
+ 0.5 nm
|
Stray Light: (B-Band for Illuminant A)
|
2.5 - 10E-4
|
Dynamic Range:
|
1000000:1
|
Linearity:
|
+ 0.8%
|
A/D Converter:
|
16 bit
|
A/D Rate:
|
1 MHz
|
PC Interface:
|
USB 2.0
|
Weight:
|
6 kg (18 lbs)
|
Dimensions: (W x H x D)
|
105 x 230 x 282 mm | CDS-9800-VISB Spectrometer (300 - 950 nm)
Detector: (*1)
|
TE Cooled 1024 x 122 CCD
|
Spectral Range:
|
300 - 950 nm |
Grating:
|
f= 85.8 mm 340 grooves/mm
|
Gate-Time:
|
5 msec - 20 msec (option: 5 msec to 65 sec)
|
ND Filter:
|
OD1-2
|
Data Point Interval:
|
0.7 nm (1024 ch)
|
Resolution: (nm/ch)
|
1.4 nm (512 ch)
|
Wavelength Accuracy:
|
+ 0.3 nm
|
Stray Light: (B-Band for Illuminant A)
|
2.5 - 10E-4
|
Dynamic Range:
|
1000000:1
|
Linearity:
|
+ 0.8%
|
A/D Converter:
|
16 bit
|
A/D Rate:
|
1 MHz
|
PC Interface:
|
USB 2.0
|
Weight:
|
6 kg (18 lbs)
|
Dimensions: (W x H x D)
|
105 x 230 x 282 mm | CDS-9800-UV-VIS Spectrometer (240 - 800 nm)
Detector: (*1)
|
TE Cooled 1024 x 122 CCD
|
Spectral Range:
|
240 - 800 nm
|
Grating:
|
f= 85.8 mm 405 grooves/mm
|
Gate-Time:
|
5 msec - 20 msec (option: 5 msec to 65 sec)
|
ND Filter:
|
OD1-2
|
Data Point Interval:
|
0.6 nm (1024 ch)
|
Resolution: (nm/ch)
|
1.2 nm (512 ch)
|
Wavelength Accuracy:
|
+ 0.3 nm
|
Stray Light: (B-Band for Illuminant A)
|
2.5 - 10E-4
|
Dynamic Range:
|
1000000:1
|
Linearity:
|
+ 0.8%
|
A/D Converter:
|
16 bit
|
A/D Rate:
|
1 MHz
|
PC Interface:
|
USB 2.0
|
Weight:
|
6 kg (18 lbs)
|
Dimensions: (W x H x D)
|
105 x 230 x 282 mm |
|